Magnetic resonance imaging with spatially-selective pulses using multiple transmission channels

by Denis Kokorin




Institution: Universität Freiburg
Department: Mathematik und Physik
Degree: PhD
Year: 2015
Record ID: 1115606
Full text PDF: http://www.freidok.uni-freiburg.de/volltexte/2015/9950/


Abstract

This work explores the use of spatially-selective RF excitation combined with parallel transmission for inner volume MRI applications in humans. The technique of parallel excitation allows to confine MR imaging to an arbitrarily-defined region of interest selected in the scanned subject. This principle offers numerous advantages and was tested in phantoms and human subjects on two MRI systems, which were combined with a special hardware extension with 8 independent RF transmit channels. Advantages and disadvantages of the implemented technique were evaluated and analyzed based on the experimental data obtained. In dieser Arbeit wurden die Techniken der räumlich-selektiven HF-Anregung in Kombination mit dem parallelen Senden für MRT-Anwendungen in Menschen analysiert. Mit Hilfe der parallelen Anregung ist es möglich, die MR-Bildgebung auf beliebige kleine Bereichen im zu untersuchten Objekten zu beschränken. Diese Technik bietet zahlreiche Vorteile für die MRT und wurde sowohl in Phantomen als auch in Probanden mit zwei verschiedenen MRT-Systemen getestet. Die beiden MR-Systeme hatten spezielle Apparaturen für das parallele Senden mit acht unabhängigen HF-Kanälen. Die Vorteile und Nachteile der realisierten MR-Methoden wurden anhand der experimentellen Ergebnisse herausgearbeitet.